Rosemount / Emerson
Rosemount 3051 Pressure Transmitter
A configurable transmitter family for pressure measurement and pressure-derived level and flow applications.
Product Overview
The Rosemount 3051 is a configurable pressure-transmitter family. It directly measures pressure; selected configurations can support level or flow measurement methods. Individual model selection remains required.
Key Features
- Up to 0.04% reference accuracy and up to 150:1 rangedown, depending on the selected configuration.
- Available graphical back-lit display and Bluetooth® connectivity for supported configurations.
- Available Process Alerts, Loop Integrity, and Plugged Impulse Line diagnostics.
- Available wired, low-power, fieldbus, PROFIBUS®, and WirelessHART® communication options.
Technical Specifications
| Measurement roles | Pressure; pressure-derived level and flow configurations |
|---|---|
| Reference accuracy | Up to 0.04% (configuration-specific) |
| Rangedown | Up to 150:1 (configuration-specific) |
| Communication | 4–20 mA HART®, WirelessHART®, FOUNDATION™ Fieldbus, PROFIBUS®, 1–5 V Low Power HART® |
| Measurement range | Coplanar: up to 4,000 psi; inline: up to 20,000 psi (configuration-specific) |
| Process-wetted materials | Available options include 316L SST, Alloy C-276, Alloy 400, Tantalum, and gold-plated variants; selection depends on model configuration |
| Diagnostics | Available Process Alerts, Loop Integrity, and Plugged Impulse Line diagnostics |
Available Configurations
- Coplanar
- Inline
- Level
- Annubar
- Compact Orifice
- Integral Orifice
- Hygienic
These are product-family configuration paths. Final suitability, limits, materials, approvals, and diagnostics must be confirmed against the selected model and official documentation.
Industrial Applications
- Pressure measurement
- Hydrostatic and differential-pressure level measurement
- Differential-pressure flow measurement with the appropriate primary element
Application suitability depends on the selected transmitter, process connection, measurement range, materials, and installed assembly.
Official Technical Documents
RFQ Guidance
For an accurate review, provide the complete model code or nameplate when available, the measured variable, pressure range, process connection, communication requirement, materials or service conditions, quantity, and any supporting documentation.